1. Remove the tip ejector by holding down the ejector and inserting the tool supplied to relese the locking mechanism carefully,release the "tip ejector" and "shaft". |
2. Slowly turn anti-clockwise "tip cone". |
3. Remove the tip ejector by holding down the ejector and inserting the tool supplied to relese the locking mechanism carefully,release the "tip ejector" and "shaft". |
4. Placethe"tipcone"
,"tipejector","shaft",
"pistone","o-ring"
and "spring" into
the beaker
containing 70%
ethyle alchole
and keave
for atleast
30mints. |
5. After performing the above procedure remove the components from the beaker and rinse them with distilled water,then dry preferably with warm air,for atleast an hour. |
6. Re-grease the piston as described in tje instruction manual. |
Note: Every 6 month when you performing Decontimination procedure kindly check ' 0 ' -ring and if possible kinly replaced if necessary.The calibration has to be checked after cleaning the pipette. |