| TROUBLE |
POSSIBLE REASON |
SOLUTION |
| Drop left inside the Tip |
Unsuitable Tip |
Use new "Biotip" |
| Aspiration Problem |
Fitting of tip cone is loose (or) |
Screw the tip cone firmly to the body |
| Wrong placing of ''O ring'' |
Refer ''CD Manual Guide'' Reset the rings accordingly |
| Leakage Problem |
Tip incorrectly attached |
Attach firmly |
| Foreign particles between tip & tip cone |
Clean tip cones attach new tips |
Foreign particles between seal, O-ring, piston and tip
cone |
Clean seal, O-ring, Piston & tip cone with napless
cloth |
| Insufficient amount of grease on piston & O ring |
Grease accordingly |
| O-ring not correctly positioned or damaged |
Change the O-ring |
| Tip cone Scratched or damaged |
Change the tip cone |
| Inaccurate dispensing |
Incorrect operation |
follow instructions carefully |
| Tip incorrectly attached> |
Attach firmly |
| Calibration altered |
according to instruction |
| Instrument damaged |
Send for repair |
| Push button Jammed |
Piston contaminated |
Clean and grease O-ring and piston. |
| Penetration of solvent fumes |
| Tip Ejector jammedor shaft broken |
Tip ejector spring is loose |
Replace the spring and spring support. |
| |
Tip cone contaminated from outside |
Remove tip ejector shaftclean tip cone outer surface with ethanol. |
| |
Due to tip ejector pressure or misused |
Replace the shaft |
| Volume Setting is not properly click or over rotate |
Mechanism damaged |
Send for repair |